Project Description

PT-LC1100 Isocratic/binary gradient

High Pressure Constant Flow Pump

  •  Double plunger reciprocating pump
  •  Flow Range:0.001mL/min-10.000mL/min;
  • Setting Step Size:0.001mL/min
  •  Flow Accuracy:≤±0.3% (1.00ml/min, H2O,R.T.,8.5MPa)
  •  Flow stability:RSD≤0.1% (1.00ml/min, H2O,R.T.,8.5MPa)
  •  Maximum Pressure:42MPa
  •  Display Pressure Error:≤±3.0%
  • Pressure Fluctuation:≤1.0%
  •  Power Dissipation:80W
  •  External Dimension
  • (Length x Width x High):420mm×260mm×160mm

UV/VIS Detector

  • Wavelength Range:190-700nm
  • Wavelength Repeatability:±0.1nm
  • Wavelength Accuracy:≤±1.0nm
  • Spectrum Band Width:8nm
  • Response Time:0.0s-4.9s
  • Linearity Range:1.8AU (5%)
  • Baseline Noise:≤±1×10-5AU
  • Baseline Drife:≤2.0×10-4AU/h
  • minimum Detectable Concentration:4.0×10-9g/mL (Naphthalene in Methanol Solution)
  • Detector Cell:Optical Path–10mm,Volume–10μL
  • Light Source:Deuterium Lamp
  • Power Dissipation:100W
  • External Dimension
  • (Length x Width x High):420mm×260mm×160mm