Project Description
HPLC PT-LC3100
- High Pressure Constant Flow Pump for Isocratic and Binary gradient
- Flow rate range: 0.001-10.000 mL/min, in 0.001 mL increments
- Flow accuracy: ≤±0.2% (1.00ml/min, H2O,R.T.,8.5MPa)
- Flow precision: RSD≤0.075% (1.00ml/min, H2O,R.T.,8.5MPa)
- Maximum pressure: 45 MPa
- Pressure accuracy: ≤±3.0%
- Pressure pulsation: ≤0.1 MPa
Low-Pressure Gradient Unit for Quaternary
- Integrated vacuum degasser, 4 channel
- Composition rang: 0-100%
- Composition precision: ≤0.2% RSD
- Composition accuracy: ≤±2.0%
- UV detector
- Wavelength range: 190-700 nm
- Wavelength repeatability: ≤0.1 nm
- Wavelength accuracy: ≤±1.0 nm
- Bandwidth: 8nm
- Response time: 0.0 s-5 s
- Baseline noise: ≤±0.5×10-5 AU
- Drift: ≤1.0×10-4 AU/h (empty cell, 254 nm, 60 min)
- Path length: 10 mm(analytical cell)
- Light source: Deuterium Lamp
- Refractive Index Detector
- Flow cell type:2 chamber-type
- Measuring method:Deflection type
- Refractive Index range:00 to 1.75
- Measuring range:25 to 512μRIU
- Drift:2μRIU/h (Pure water 1mL/min, PURGE OFF)
- Linearity range:≥600μRIU
- Noise:≤2.5nRIU (Pure water, response : 1.5sec)
- Response:1 , 0.25 , 0.5 , 1 , 1.5 , 2 , 3 , 6 sec
- Cell volume:8μL
- Flow rate:(Usual) 0.2 to 3.0mL/min;(Max.) 10mL/min (solvent: pure water)
- Evaporative Light Scattering Detector
- Light source: high intensity halogen lamp
- Detector: high sensitivity photomultiplier
- Analogue signal output: 0-1V
- Heating time:<10min heat-up to 150℃
- Temperature range: evaporation chamber max 150℃, increase by 1℃, nebulizer max 70℃
- Gas flow rate: 1-2L/min, recommended pressure 1-2bars
- Noise<1m Volt(1ml/min)
- Nebulizer type: venture tube
- Diode Array Detector
- 512 diodes
- Wavelength range: 190-760 nm
- Wavelength repeatability: ±0.1 nm
- Wavelength accuracy: ±1nm
- Baseline noise: ≤±1.0×10-5 AU (empty cell, 254 nm, 1.0 s)
- Drift: ≤2.0×10-4 AU/h (empty cell, 254 nm, 60 min)
- Path length: 10 mm (analytical cell)
- Light source: Deuterium Lamp and Tungsten Lamp
- Fluorescence Detector
- Light source: Xenon flash lamp
- Wavelength Range: 0, 200 nm to 650 nm (Dual wavelength simultaneous monitoring possible)
- Spectral Bandwidth: 20 nm
- Wavelength Accuracy: ±2 nm
- Wavelength Reproducibility: ±0.2 nm
- S/N: Water Raman peak S/N 1200 min
- Cell (capacity, pressure resistance, material): 12 µL; 2 MPa (approx. 20 kgf/cm2); SUS316L, PTFE (fluororesin), quartz
- Autosampler
- Inject volumn range: 0-100 μL (standard)
- Precision: RSD <0.3%
- linearity: >0.999
- Sample residual: <0.01%
- Number of sample plates: Two of 60 position 2-mL vial plates
- Syringe: 500 μL
- Column oven
- Column capacity: Three columns, up to 10mm internal diameter, up to
- 250mm length with guard column
- Temperature range: room temperature+5.0 to 80℃, settable in 0.1℃ increments
- Temperature accuracy: ≤±0.5℃
- Temperature stability: ≤±0.1℃
Optional column oven(cooling function)
- Column capacity: 3 columns up to 300mm length
- Cooling function temperature range: 5 to 80℃
- Normal temperature range: room temperature to 100℃
- Accuracy: 0.1℃
- Displayer: 4-digit LED
- Workstation
- Internal and external standard calculation methods
- Automatically displays, prints, exports and starts other programs after the completion of a measurement.
- Comply with the requirements of the 21 CFR Part 11 directive of the FDA
- On-line degasser
- Maximum number of degassing Channel: 4
- Flow path ID: 1.143mm
- Internal volume: 480μL
- Wetted Materials: PEEK, Teflon® AF